An Eco-friendly RP-HPLC and UV-Method Development and Validation for an Estimation of Tolvaptan in Bulk and Tablet Dosage form Followed by Forced Degradation Studies
Patel Seema A. *
Department of Quality Assurance, Ali-allana College of Pharmacy, Akkalkuwa, Nandurbar, Maharashtra, 425 415-India.
Sayyed Nazifa S.
Department of Quality Assurance, Ali-allana College of Pharmacy, Akkalkuwa, Nandurbar, Maharashtra, 425 415-India.
Lajporiya Mobina I.
Department of Quality Assurance, Ali-allana College of Pharmacy, Akkalkuwa, Nandurbar, Maharashtra, 425 415-India.
Manjra Mehfuza U.
Department of Quality Assurance, Ali-allana College of Pharmacy, Akkalkuwa, Nandurbar, Maharashtra, 425 415-India.
Aejaz Ahmed
Department of Quality Assurance, Ali-allana College of Pharmacy, Akkalkuwa, Nandurbar, Maharashtra, 425 415-India.
G. J. Khan
Department of Quality Assurance, Ali-allana College of Pharmacy, Akkalkuwa, Nandurbar, Maharashtra, 425 415-India.
*Author to whom correspondence should be addressed.
Abstract
Aims: To develop and validate a new, simple, rapid, precise, and accurate An Eco-friendly RP-HPLC and UV-Method Development and Validation for an estimation of Tolvaptan in Bulk and Tablet dosage form followed by Forced Degradation Studies
Place and Duration of the Study: The present work has been carried out at Ali-Allana College of Pharmacy, Akkalkuwa between November-2020 to April-2021.
Methodology: The UV-Spectroscopic method was developed for the estimation of tolvaptan in bulk and tablet dosage form. The solvent selected for the tolvaptan UV analysis was 4% aq. SLS solution, the solution of 10µg/ml was scanned in UV region from 200-400 nm and the λmax value was determined. The RP-HPLC method was developed on Sunsil C18 150 mm x 4.6mm x 5μ column using acetonitrile: water [45:55] as mobile phase at flow rate 1.0 ml/min and UV detection at 266 nm.
Results: The maximum absorbance was observed at 266 nm. The wavelength 266 nm was selected for further analysis of tolvaptan. The calibration curve was determined using drug concentrations ranging from 20-100 µgm/ml. The system suitability was performed by injecting a standard solution containing 200µg/ml of tolvaptan in six replicates. For two of them, the peak asymmetric were <1.5 and the theoretical plate number was >2000, and the %RSD of tolvaptan was less than 2.
Conclusion: From the above results, it was concluded that the developed UV and RP-HPLC methods are precise and accurate and can be applied for the quantitative estimation of tolvaptan from bulk and tablet dosage forms. The method can be used for routine testing of tolvaptan by the pharmaceutical industry. Validation of the developed method was done as per International Conference on Harmonization (ICH) Q2R1 guidelines.
Keywords: Reverse phase high-performance liquid chromatography, UV-visible spectroscopy, method development, validation, tolvaptan, international conference on harmonization (ICH) Q2R1